Clinical aid
Published on 03/09/96
Questions and answers
By Alberto Mazzocchi MD, DDS
According to Starnbach(1)
orthodontists have the second highest incidence of Hepatitis B among dental workers.
Sterilization of orthodontic pliers is necessary to avoid any risk of cross infection.(2-3-4)
Although unlike surgeons, orthodontists generally do not work in a blood contaminated
area, orthodontic arch wires and ligatures can traumatize patients' mucosa, causing
bleeding.
A recent survey indicates that orthodontists and chairside assistants receives an average
of about one cut per week.
Orthodontic instruments present special
problems since they have large, difficult to clean hinge areas, sharp angles, cutting
edges or pointed ends that can be potentially damaged by corrosion using autoclave
sterilizers which use water vapors.
Chemiclave (chemical vapors) sterilizers can be used with minimal corrosion and damage to
cutting edges.
Rapid Dry Heat (RDH) sterilizers which use no vapors have no potential risk of corroding
instruments.
Chrome plated pliers are more resistant to
damage during sterilization cycles than stainless steel pliers (4).
A good way to test metal quality of orthodontic pliers is performing microhardness tests.
An improvement in hardness following sterilization procedures can be explained as an
increase in brittleness and a decrease in corrosion resistance.
There is not much data available on microhardness tests. Mazzocchi et Coll (6)
reported a study on Ormco/AEZ and 3M-Unitek (old series) orthodontic cutters in 1994.
New unpublished tests are shown at the end of this article (Table 1).
To avoid corrosion of pliers, manufacturers recommend the following procedures:
Yes. The OSHA and EPA requirements for
sterilization procedures forced manufacturers to produce instruments which could undergo
several sterilization cycles.
This has translated into better quality instruments being offered to customers during
recent years.
For example the 3M-Unitek Prestige Series showed a high improvement in metal hardness in
comparison to the previous series.
Yes. AEZ/Ormco pliers have a one-year
warranty against corrosion and damage.
3M-Unitek Prestige Series pliers have a five-year warranty against corrosion and damage.
Dentronix pliers have a lifetime perpetual warranty.
Chemiclave and Rapid Dry Heat sterilizers
are the most useful in preventing corrosion. (4-5)
Chemiclave, which uses an alcoholic solution, reduces rust on instruments, has a short
time of sterilization and a low temperature in the sterilization cycle (270°F).
Instruments may be inserted in polyethylene paper pouches and stored in the office.
The major disadvantages are chemical solution costs and fumes.
Rapid Dry Heat units were proposed at the end of 80's, in order to avoid staining and
fumes and to reduce time of sterilization using a higher temperature (340°F).
Instruments can be positioned on the proper racks without packaging.
Major disadvantages are electricity costs and light color changes (browning) of the
pliers metal surface.
A study on the effects of these two sterilization methods on orthodontic pliers (3M-Unitek
and Ormco/AEZ) showed that RDH seems to produce minimal effects on metal hardness even if
these results are less predictable than chemiclave ones (Table 1).
Yes. Lubricating hinges after every
sterilization cycle is very important to preserve pliers.
Silicone lubricants (AEZ # 803-0701) are better than oil lubricants.
Silicone can be used to lubricate hinges even before sterilization.
Oils should be avoided because of the risk of interfering with heat conduction during
sterilization.
Yes. Orthodontic instruments must be
arranged in appropriate trays, packages or sterilization pouches to facilitate a well
organized cross infection control program.
Everything that will not be used immediately should be wrapped or packaged in order to
keep it sterile.
Covered metal trays are very popular in the UK (4).
Since aluminum appears to be an environmental pollutant, IMS cassettes have become more
popular during the last few years.
They are light weight resin cassettes which have a capacity for 12-18 instruments and a
section to hold disposable or miscellaneous items .(4)
Trays provide safe and aseptic storage of instruments with no risk of being punctured by
sharp instruments.
Sterilization pouches are very useful for single instrument storage.
They enable easy identification of the instruments and may be stored safely for one year.
2 groups of 4 orthodontic pliers received 400 sterilization cycles:
Each group of cutters contained:
The cutters were inserted into the
recommended polyethylene-paper pouches, as recommended by manufacturers, for chemiclave
cycles.
For RDH sterilization, pliers were positioned on the proper racks without packing them.
Every 100 cycles, pliers were cleaned and lubricated.
Orthodontic cutters were checked for Vickers hardness, discoloration and tarnish, before
and after sterilization.
Each cutter was measured with a Durimet Microhardness Tester.
The tests were performed using a force of 500g on the cutting edges (4 tests for each
cutter).
A mean value of the 4 tests was finally reported.
Discoloration and tarnish were judged by 2 metallurgic experts (not orthodontists) without
SEM or lens. Grading was defined as:
Ligature cutters microhardness before and after 400
sterilization cycles
a : 3M Unitek pin and ligature cutter
No.753
b : Eczurra one piece cutter No. 0512
| before | after | VAR | before | after | VAR | ||
|---|---|---|---|---|---|---|---|
| Vickers a | 814.25 | 878 | +64 (+7.8%) | 905 | 891 | -14 (-1.5%) | |
| Vickers b | 676 | 758 | +82 (+12.1%) | 565 | 705 | +140 (+24.7%) | |
| discoloration a | 0 | 0 | 0 | 0 | 0 | 0 | |
| discoloration b | 0 | 0 | 0 | 0 | 0 | 0 | |
| tarnish a | 0 | 0 | 0 | 0 | 0 | 0 | |
| tarnish b | 0 | 0 | 0 | 0 | 0 | 0 |
c : 3M Unitek safety hold universal
cutter No. 756
d : AEZ/Ormco minidistal-end cutter with hold No. 0501
| before | after | VAR | before | after | VAR | ||
|---|---|---|---|---|---|---|---|
| Vickers c | 742.25 | 791 | +49 (+6.6%) | 757 | 804 | +47 (+6.2%) | |
| Vickers d | 774 | 817 | ++43 (+5.5%) | 699 | 692 | -7 (-1.0%) | |
| discoloration c | 0 | 0 | 0 | 0 | 0 | 0 | |
| discoloration d | 0 | 0 | 0 | 0 | 0 | 0 | |
| tarnish c | 0 | 0 | 0 | 0 | 0 | 0 | |
| tarnish d | 0 | + | + | 0 | 0 | 0 |
Table 1- The effects of 400 sterilization cycles on AEZ/Ormco and 3M
UNITEK-Prestige Series cutters
(Mazzocchi AR, Caprioglio A. 1996)